Z-1581-2023 Class II Ongoing

Recalled by Smiths Medical ASD Inc. — Minneapolis, MN

FDA device recall Z-1581-2023 was initiated by Smiths Medical ASD Inc. on Apr 5, 2023 and is designated Class II. Reason for recall: DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of… The recall status is ongoing. Affected quantity: 2101 units.

Recall details

Product Type
Devices
Report Date
May 24, 2023
Initiation Date
Apr 5, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2101 units

Product description

DuraLife Autoclavable Silicone Extension Tube, REF 60-1502 (UltraSet Product Code 66-2509); Non-Ported Double Swivel Elbow for use with breathing circuits

Reason for recall

DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a single patient.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, New Zealand, and Hong Kong.

Code information

3569291, 3802463, 3974741, 4093611, 4219069, 4340640, 3582651, 3832622, 4001787, 4108514, 4242510, 4340642, 3609721, 3871182, 4011956, 4116120, 4254340, 3609743, 3891814, 4032896, 4122394, 4264898, 3693349, 3914797, 4041943, 4151071, 4279831, 3723464, 3927549, 4063256, 4167796, 4290272, 3761125, 3952707, 4090324, 4182390, 4340638

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1581-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Smiths Medical ASD Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Smiths Medical ASD Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Smiths Medical ASD Inc. →

Source: openFDA record for Z-1581-2023. Report a data issue: hello@regdatalab.com.