Z-1587-2016 Class II Terminated

Recalled by Livongo Health — Chicago, IL

FDA device recall Z-1587-2016 was initiated by Livongo Health on Feb 18, 2016 and is designated Class II. Reason for recall: The instructional insert for the In Touch Control Solution for use with the In Touch meter and In Touch strips as a quality control check did not contain the limitation of sensor accuracy: (1)… The recall status is terminated (terminated Mar 6, 2018). Affected quantity: 63,000 units.

Recall details

Product Type
Devices
Report Date
May 11, 2016
Initiation Date
Feb 18, 2016
Termination Date
Mar 6, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
63,000 units

Product description

The Livongo Health In Touch¿ Blood Glucose Monitoring System (BGMS), is an Over-The-Counter (OTC) system designed for the self- monitoring of blood glucose by persons with diabetes in home settings to aid in their diabetes management.

Reason for recall

The instructional insert for the In Touch Control Solution for use with the In Touch meter and In Touch strips as a quality control check did not contain the limitation of sensor accuracy: (1) with the use of acetaminophen (Tylenol) (2) with the use of Ascorbic Acid (vitamin C) supplementation (3) in the presence of uric acid

Distribution pattern

US consignees: Arizona, California, Delaware, Florida, Georgia, Illinois, Kentucky, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia and Washington.

Code information

Model/Reorder Number 3614-001-001 with the following lot numbers 890071 890072, 890073, 890074, 890075, 890076, 890077, 890078 and 890079.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1587-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Livongo Health? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Livongo Health: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Livongo Health →

Source: openFDA record for Z-1587-2016. Report a data issue: hello@regdatalab.com.