Z-1602-2024 Class II Ongoing

Recalled by Abbott Molecular, Inc. — Des Plaines, IL

Recall Details

Product Type
Devices
Report Date
May 1, 2024
Initiation Date
March 20, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,012 units

Product Description

Alinity m System, Part Number: 08N53-002

Reason for Recall

The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMBODIA, CAMEROON, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KENYA, LATVIA, LUXEMBOURG, MALAWI, MALAYSIA, MEXICO, MONTENEGRO, MOROCCO, MOZAMBIQUE, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PAKISTAN, PANAMA, PARAGUAY, POLAND, PORTUGAL, ROMANIA, RUSSIA (CIS), SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TURKEY, UGANDA, UNITED KINGDOM, VIETNAM, ZIMBABWE.

Code Information

Part Number: 08N53-002; UD/DI: 00884999048034. Not Serial Number specific