Z-1614-2016 Class I Terminated

Recalled by Monteris Medical Corp — Plymouth, MN

FDA device recall Z-1614-2016 was initiated by Monteris Medical Corp on Mar 25, 2016 and is designated Class I. Reason for recall: Monteris received a report that the sapphire lens on a 2.2SF probe fractured during a LITT procedure, with resulting introduction of C02 into the patient's brain. The recall status is terminated (terminated Mar 10, 2017). Affected quantity: 121.

Recall details

Product Type
Devices
Report Date
May 25, 2016
Initiation Date
Mar 25, 2016
Termination Date
Mar 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
121

Product description

NeuroBlate System SideFire Select 2.2 mm Directional Laser Probe, Sterile EO. Model #s SFS000-01,SFS002-01, SFS122-01, SFS222-01, SFS322-01, SFS422-01, and SFS522-01. NeuroBlate System SideFire Select Directional Laser Probes are part of the NeuroBlate System. The Monteris Medical NeuroBlate System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It also provides real-time thermographic analysis of selected MRI images. The laser delivery probes direct the laser energy to the proper treatment location.

Reason for recall

Monteris received a report that the sapphire lens on a 2.2SF probe fractured during a LITT procedure, with resulting introduction of C02 into the patient's brain.

Distribution pattern

Nationwide Distribution in the states of KS, MO, NC, OH, and VA.

Code information

All unexpired lots of these models are being removed from the market.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1614-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Monteris Medical Corp? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Monteris Medical Corp: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Monteris Medical Corp →

Source: openFDA record for Z-1614-2016. Report a data issue: hello@regdatalab.com.