Z-1615-2024 Class II Ongoing

Recalled by XENEX Disinfection Services Inc. — San Antonio, TX

Recall Details

Product Type
Devices
Report Date
May 1, 2024
Initiation Date
March 18, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
377

Product Description

LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the LightStrike Germ-Zapping Robots Models: PXUV4D, PXUV5D, LSMUV6-FT, or LSMUV6-SL, for whole room microbial reduction.

Reason for Recall

Blackout curtains are no longer recommended as a protective barrier against UV/Ozone exposure to a bystander standing in the same room, but on the other side of the curtain from where a microbial reduction device UV cycle is taking place. Bystanders in the room during the cycle may experience temporary skin reddening, feeling of sand in eyes, nausea, headache, and/or mucous membrane irritation.

Distribution Pattern

US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, ID, IL, KY, LA, MI, MN, MO, NC, ND, NV, OH, OK, OR, TN, UT, WA, WV, WY.

Code Information

Robot UDI-DIs: 00814122020033, 00814122020064, 00814122020125, 00814122020132. All serial numbers.