Z-1639-2022 Class II

Recalled by Aggredyne, Inc. — Houston, TX

FDA device recall Z-1639-2022 was initiated by Aggredyne, Inc. on Jul 23, 2019 and is designated Class II. Reason for recall: The firm identified a software design issue where the improper shut down of the instrument resulted in speeds.csv file contents of the software version 5.1 being erased. Due to this issue, the… Affected quantity: 47.

Recall details

Product Type
Devices
Report Date
Sep 7, 2022
Initiation Date
Jul 23, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47

Product description

AggreGuide A-100 Instrument

Reason for recall

The firm identified a software design issue where the improper shut down of the instrument resulted in speeds.csv file contents of the software version 5.1 being erased. Due to this issue, the target RPM for the assay is 0 and the motor output required to achieve this RPM is also 0. Therefore, when the assay is run, the software applies 0 Volts to the motor and performs a 0 RPM assay. Because of this, there is no mixing of the blood with agonist and no moving of aggregates happens in front of the optical detectors. The resulting PAI is 0 and the result is displayed as Low.

Distribution pattern

International distribution to the country of South Korea.

Code information

UDI-DI: 00856893004065; Software Versions 5.1 or 5.2; Serial Numbers Between xx0001 and xx0160

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1639-2022
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Aggredyne, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1639-2022. Report a data issue: hello@regdatalab.com.