Z-1641-2018 Class II Terminated

Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

FDA device recall Z-1641-2018 was initiated by Medtronic Sofamor Danek USA Inc on Mar 13, 2018 and is designated Class II. Reason for recall: The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill… The recall status is terminated (terminated Nov 2, 2020). Affected quantity: 231 units.

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
Mar 13, 2018
Termination Date
Nov 2, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
231 units

Product Description

Medtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE LP reusable instruments are intended for the surgical implantation of the PRESTIGE LP Cervical Disc. Medtronic reusable instruments must be sterilized prior to initial use and must be cleaned and sterilized prior to each re-use.

Reason for Recall

The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.

Distribution Pattern

US Nationwide Distribution

Code Information

a) REF 6975205, 5mm; UDI 00643169752979 Lot Number ID17K018 b) REF 6975206, 6mm; UDI 00643169752962 Lot Number ID17J009 c) REF 6975207, 7mm; UDI 00643169752955 Lot Numbers: ID17K007, ID17L035

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1641-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.