Z-1641-2026 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

FDA device recall Z-1641-2026 was initiated by Olympus Corporation of the Americas on February 25, 2026 and is designated Class II. Reason for recall: Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking d… The recall status is ongoing. Affected quantity: 2648 units.

Recall Details

Product Type
Devices
Report Date
April 1, 2026
Initiation Date
February 25, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2648 units

Product Description

Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electrosurgical energy under laparoscopic visualization. The device is intended to be used with the Gyrus ACMI G400 Workstation only.

Reason for Recall

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Distribution Pattern

Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.

Code Information

Model: HACF0533; UDI: 00821925036390; Lot Numbers: All unexpired;