Z-1651-2021 Class I Ongoing

Recalled by The Biomed Guys — Clarksville, TN

Recall Details

Product Type
Devices
Report Date
June 9, 2021
Initiation Date
March 25, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22 units

Product Description

Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.

Reason for Recall

Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.

Distribution Pattern

Us distribution to the states of IN and TX.

Code Information

Serial numbers of serviced devices, serviced 09/2020-02-2021: 12833065, 13187021, 13187811, 12838069, 12982509, 12837231, 12835186, 12837972, 12710593, 12776790, 13753893, 13605919, 12832544, 15763526, 14806203, 14806612, 14807070, 13738432, 15763547, 15865626, 14806472, 12796415,