Z-1666-2023 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 14, 2023
- Initiation Date
- April 11, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 556
Product Description
StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems
Reason for Recall
In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury
Distribution Pattern
US Distribution: PR. OUS (foreign) distribution: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, Colombia COLOMBIA, C¿TE D'IVOIRE, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, LATVIA, MEXICO, MOLDAVIA, NETHERLANDS, PAKISTAN, POLAND, PORTUGAL, REPUBLIC of KOREA, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED ARAB EMIRATES
Code Information
Model/UDI-DI/Software Versions: 9735585/00763000253165, 00763000306762 (inside Kits: 00763000299927, 00763000299972), 00763000517267(inside Kits: 00763000517274, 00763000517335)/v3.1.1, 3.1.2, and 3.1.3