Z-1675-2025 Class II Ongoing

Recalled by Belmont Instrument LLC — Billerica, MA

FDA device recall Z-1675-2025 was initiated by Belmont Instrument LLC on March 31, 2025 and is designated Class II. Reason for recall: Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser. The recall status is ongoing. Affected quantity: 208 sets.

Recall Details

Product Type
Devices
Report Date
May 7, 2025
Initiation Date
March 31, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
208 sets

Product Description

Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).

Reason for Recall

Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.

Code Information

Model Number: 902-00037. UDI-DI: 00896128002213. Lot Numbers: 20241012, 20241107