FDA device recall Z-1680-2023 was initiated by DePuy Mitek, Inc., a Johnson & Johnson Co. on April 13, 2023 and is designated Class II. Reason for recall: IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the pati… The recall status is ongoing. Affected quantity: 3328 units.
Recall Details
- Product Type
- Devices
- Report Date
- June 14, 2023
- Initiation Date
- April 13, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3328 units
Product Description
DePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212722
Reason for Recall
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
Code Information
UDI-DI: 10886705002085 All lot numbers