Z-1696-2018 Class II Terminated

Recalled by Straumann Manufacturing, Inc. — Andover, MA

FDA device recall Z-1696-2018 was initiated by Straumann Manufacturing, Inc. on Mar 23, 2018 and is designated Class II. Reason for recall: A change regarding the shelf life of a raw material was implemented without an appropriate submission of data demonstrating the continued efficacy of the product. The recall status is terminated (terminated May 2, 2019). Affected quantity: 2915.

Recall details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
Mar 23, 2018
Termination Date
May 2, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2915

Product description

Straumann¿ Emdogain 0.3 ml Multipack. Enamel Matrix Derivative. Article Number 075.114W

Reason for recall

A change regarding the shelf life of a raw material was implemented without an appropriate submission of data demonstrating the continued efficacy of the product.

Distribution pattern

Distributed US Nationwide and Puerto Rico.

Code information

Lot NY938, Syringe Blister Lot MN843, Exp 11/30/2018 Lot NP738, Syringe Blister Lot NE101A, Exp 3/31/2019 Lot PG629, Syringe Blister Lot NK216, Exp 5/31/2019 Lot PY817, Syringe Blister Lot PT888C, Exp 1/31/2020

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1696-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Straumann Manufacturing, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1696-2018. Report a data issue: hello@regdatalab.com.