Z-1737-2022 Class II Terminated

Recalled by Technomed Europe — Maastricht-Airport

FDA device recall Z-1737-2022 was initiated by Technomed Europe on Aug 9, 2022 and is designated Class II. Reason for recall: Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU. The recall status is terminated (terminated Apr 11, 2024). Affected quantity: 37/25-electrode boxes.

Recall Details

Product Type
Devices
Report Date
Sep 21, 2022
Initiation Date
Aug 9, 2022
Termination Date
Apr 11, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
37/25-electrode boxes

Product Description

Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per package, sterile.

Reason for Recall

Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU.

Distribution Pattern

US Nationwide distribution in the states of CA, FL, NH, OH, and WA.

Code Information

Lot numbers 044220, exp. 1/1/2023; 045077, exp. 3/1/2023; 045443, exp. 4/1/2023; and 049285, exp. 7/1/2024; UDI-DI 08718375861530.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1737-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.