Z-1752-2013 Class II Terminated

Recalled by Baxa Corporation — Englewood, CO

FDA device recall Z-1752-2013 was initiated by Baxa Corporation on Jun 4, 2013 and is designated Class II. Reason for recall: Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the… The recall status is terminated (terminated Oct 16, 2015). Affected quantity: 753 units.

Recall details

Product Type
Devices
Report Date
Jul 24, 2013
Initiation Date
Jun 4, 2013
Termination Date
Oct 16, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
753 units

Product description

ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients into a finished solution.

Reason for recall

Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight December 31, 2010.

Distribution pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Canada, Latin America, Europe/Middle East/Africa, and Taiwan.

Code information

ExactaMix 1200 serial numbers 45004-45481 (sequentially), and ExactaMix 2400 serial numbers 40005-40335.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1752-2013
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Baxa Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1752-2013. Report a data issue: hello@regdatalab.com.