Z-1753-2020 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 29, 2020
- Initiation Date
- March 17, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2425 Units.
Product Description
RaySert PLUS RSP01. LOT K030. Expiry of 2024-01 - Product Usage: The single use injectors (Model STWOl and Model RSPOl) are intended to be used to compress and insert into the capsular bag only those intraocular lenses that allow the use of these injectors in their approved labelling.
Reason for Recall
RaySert PLUS RSP01 LOT K030 labeled with an expiry of 2024-01 contained an inner tray that was labeled with an expiry of 2015. In addition to having an expiry of 2015 the LOT number on the inner tray is identified as V001.
Distribution Pattern
US nationwide distribution in the states of OH, FL, GA, MD, SC, NJ, PR, IN, ID and MN.
Code Information
LOT K030.