Z-1766-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 8, 2026
- Initiation Date
- March 3, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 kits
Product Description
OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem
Reason for Recall
Oxoid Ltd., part of Thermo Fisher Scientific, confirmed that Microbact 12L Kit 20 Tests MB1128A lot 4494873 may identify incorrect organism due to inconclusive result from ambiguous color reactions in wells.
Distribution Pattern
US Nationwide distribution in the states of GA and CA.
Code Information
UDI/DI 5032384127498, Lot Number 4494873