Z-1767-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 8, 2026
- Initiation Date
- February 27, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 41
Product Description
SIGNA Premier systems
Reason for Recall
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of
Code Information
UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677