Z-1774-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 19, 2017
- Initiation Date
- March 28, 2017
- Termination Date
- September 17, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,156 (total for all devices)
Product Description
Kappa, Catalog Number: MS18853 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Catalog Number(s): MS18597/MS18596/MS18853. Product Usage: The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patients medical condition. The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2, which for the neonatal population, is to only be used when the patient is not under gas anesthesia. For combination with Scio gas module: Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the In-finity monitors.
Reason for Recall
It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of: AL, CA, GA, IL, LA, MA NY, OH, PA, TN, TX and VA and the countries of: AR AT BE BG BN BO BY CA CH CL CO CZ DE EG ES FI FR GB GR HR HU ID IE IL IN IT LT MX MY NL NO NP PA PE PK PL PT RO RS RU SA SE SK TH TR TW UG UZ VN ZA
Code Information
VF10.0 software