Z-1781-2013 Class II Terminated

Recalled by Myoscience Inc — Redwood City, CA

FDA device recall Z-1781-2013 was initiated by Myoscience Inc on Jun 21, 2013 and is designated Class II. Reason for recall: The outer box label of the product was labeled with a down revision label which did not include the US labeling requirements, specifically the symbol descriptors and also the "Caution: Federal Law… The recall status is terminated (terminated Jul 15, 2014). Affected quantity: 3 units.

Recall details

Product Type
Devices
Report Date
Jul 31, 2013
Initiation Date
Jun 21, 2013
Termination Date
Jul 15, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product description

Myoscience brand iovera Smart Tip, Myoscience Cryo-Touch IV, a Cryogenic Surgical device, REF Catalog Number: STA0311-10. Product Usage: The Myoscience Cryo-Touch IV (a.k.a. iovera) is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch IV (a.k.a. iovera) is not indicated for treatment of central nervous system tissue.

Reason for recall

The outer box label of the product was labeled with a down revision label which did not include the US labeling requirements, specifically the symbol descriptors and also the "Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician." This product was previously labeled for EU distribution only.

Distribution pattern

US distribution including CA and ID.

Code information

Serial Numbers: 100100, 100101, 100103

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1781-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Myoscience Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1781-2013. Report a data issue: hello@regdatalab.com.