Z-1833-2016 Class II Terminated

Recalled by Hospira Inc. — Lake Forest, IL

Recall Details

Product Type
Devices
Report Date
June 15, 2016
Initiation Date
March 25, 2016
Termination Date
March 6, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2567 Plum A+ infusion pumps.

Product Description

The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With its primary line, secondary line, and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories. Each system includes a pumping module (hereafter called the infuser) and an assortment of disposable IV sets (hereafter called a set), and optional accessories. The Plum A+ host device contains a Connectivity Engine peripheral module that provides wired Ethernet and wireless 802.11 a/b/g local area networking capabilities. This allows the Hospira Mednet networked application software to download drug libraries to the infuser and enable the autoprogramming feature.

Reason for Recall

Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an offset in proximal pressure.

Distribution Pattern

U.S. distribution : Natrionwide. Foreign :Brazil, Canada, Chile, Colombia, Ecuador, Guatemala, Mexico, Oman, Peru, Taiwan, Turkey, United Kingdom and Uruguay.

Code Information

PLum A+ devices with the following list numbers: 12391-04-25, 12391-22-05, 12391-23-05, 12391-23-15, 12391-36-07, 12391-54-05, 12391-88-01, 12391-88-02, 12391-88-03, 12391-88-04, 12391-88-05, 20792-04-25, 20792-13-09, 20792-57-15, 20792-65-01, 20792-65-02 and 20792-88-69.