Z-1834-2024 Class II Ongoing

Recalled by Macherey Nagel Gmbh & Co. Kg — Duren, N/A

FDA device recall Z-1834-2024 was initiated by Macherey Nagel Gmbh & Co. Kg on Apr 4, 2024 and is designated Class II. Reason for recall: The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI)… The recall status is ongoing. Affected quantity: 5,200 packages.

Recall Details

Product Type
Devices
Report Date
May 22, 2024
Initiation Date
Apr 4, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,200 packages

Product Description

Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.

Reason for Recall

The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec¿ 11-Way, #102-2285, LOT 65402.

Distribution Pattern

US Nationwide distribution in the states of CO, FL, IN, NV, TX.

Code Information

Product #102-2285, UDI/DI: 00304040001479, LOT 65402.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1834-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.