Z-1834-2024 Class II Ongoing

Recalled by Macherey Nagel Gmbh & Co. Kg — Duren, N/A

Recall Details

Product Type
Devices
Report Date
May 22, 2024
Initiation Date
April 4, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,200 packages

Product Description

Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.

Reason for Recall

The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec¿ 11-Way, #102-2285, LOT 65402.

Distribution Pattern

US Nationwide distribution in the states of CO, FL, IN, NV, TX.

Code Information

Product #102-2285, UDI/DI: 00304040001479, LOT 65402.