Z-1834-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 22, 2024
- Initiation Date
- April 4, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5,200 packages
Product Description
Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.
Reason for Recall
The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec¿ 11-Way, #102-2285, LOT 65402.
Distribution Pattern
US Nationwide distribution in the states of CO, FL, IN, NV, TX.
Code Information
Product #102-2285, UDI/DI: 00304040001479, LOT 65402.