Z-1839-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 16, 2021
- Initiation Date
- April 30, 2021
- Termination Date
- February 5, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 439 packs
Product Description
NICO Myriad-LX Illumination Pack: REF: NN-1000 Mvriad-LX Illumination Pack. 11ga REF: NN-1001 Mvriad-LX Illumination Pack. l3ga Intended to perform resection and removal of soft tissues and fluids under direct visualization accessories to the NICO Myriad System.
Reason for Recall
(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the patient or end user.
Distribution Pattern
US Nationwide Distribution
Code Information
Lot Numbers: 37KII300 37KI1301 37LI0320 37LI0322 37AJ1438 37AJ1439 37BJ0513 37BJ0516 37CJ0203 37CJ0204 37GJ1603 37GJ1607 37110208 37IJ0209