Z-1869-2025 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 18, 2025
- Initiation Date
- May 12, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14,280 units (724 cases)
Product Description
IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.
Reason for Recall
Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
Distribution Pattern
US distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX, VA, WA & WI.
Code Information