Z-1876-2024 Class I Ongoing

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

FDA device recall Z-1876-2024 was initiated by MEDLINE INDUSTRIES, LP - Northfield on February 26, 2024 and is designated Class I. Reason for recall: Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were al… The recall status is ongoing. Affected quantity: 168,632 tubes and 13,092 kits.

Recall Details

Product Type
Devices
Report Date
June 5, 2024
Initiation Date
February 26, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
168,632 tubes and 13,092 kits

Product Description

Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF DYNJ18865, Polyurethane Cuff 6.5 mm ID; (3) REF DYNJ18870, Polyurethane Cuff 7.0 mm ID; (4) REF DYNJ18875, Polyurethane Cuff 7.5 mm ID; (5) REF DYNJ18880, Polyurethane Cuff 8.0 mm ID; and (6) REF DYNJ18885, Polyurethane Cuff 8.5 mm ID. Medline SubG Endotracheal Tube with Subglottic Suctioning packaged into the following convenience kits: (1) Medline Intubation Kit - Drawer 3, Reorder No. ACC010502, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 6 kits/case; (2) Medline Intubation Tube - Drawer 3, Reorder No. ACC010527, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 6 kits/case; (3) Medline Adult Intubation, Reorder No. ACC010540, containing 8.5 size subglottic cuff, 3 kits/case; (4) Medline Adult Intubation, Reorder No. ACC010540A, containing 8.5 size subglottic cuff, 3 kits/case; (5) Medline Airway Kit - Drawer 3, Reorder No. ACC010717, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 3 kits/case; (6) Medline Airway Kit, Reorder No. DYNDJ1132, containing 7.0, 7.5, and 8.0 sizes subglottic cuff, 26 kits/case; (7) Medline Adult Intubation W Evac, Reorder No. DYNDJ1133, containing 7.0, 7.5, and 8.0 sizes subglottic cuff, 10 kits/case; (8) Medline Backup Kit, Reorder No. DYNJ909501, containing 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 1 kit/package; (9) Medline Backup Kit, Reorder No. DYNJ909501A, containing 6.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 1 kit/package; (10) Medline Adult Intubation Tray, Reorder No. DYNJAA269, containing 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 10 kits/case.

Reason for Recall

Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the suction pump is difficult to connect or detaches during use.

Distribution Pattern

Distribution was nationwide. There was government/military distributon. Foreign distribution was made to San Salvador and United Arab Emirates.

Code Information

UDI-DI numbers for the individual tubes: (1) DYNJ18860 - UDI-DI 20888277652672; (2) DYNJ18865 - UDI-DI 20888277652689; (3) DYNJ18870 - UDI-DI 20888277652696; (4) DYNJ18875 - UDI-DI 20888277652702; (5) DYNJ18880 - UDI-DI 20888277652719; (6) DYNJ18885 - UDI-DI 20888277657318. Kit code information: (1) Kit ACC010502 - Kit UDI-DI: 10193489846171; Case UDI-DI: 40193489846172; lot numbers: 23FDC233, 23FDB211, 23EDC153, 23EDB775, 23DDB280, 23CDB943, 23CDA556, 22LDA029, 22IDA550, 22HDA813, 22FDC033, 22CDC284, 21JDC054, 21IDB345, 21HDB405, and 21DDB305; (2) Kit ACC010527 - Kit UDI-DI: 10193489884333; Case UDI-DI: 40193489884334; lot numbers: 23JDA443, 23IDA363, 23IDA232, 23HDC331, 23HDA117, 23GDB600, 23FDC372, 23FDB913, 23FDA966, 23FDA310, 23EDC371, 23EDC154, 23EDC094, 23EDB732, 23EDA387, 23CDB838, 23CDA557, 23BDC008, 23BDA945, 23ADB791, 22LDA005, 22JDB800, 22HDB976, 22HDA451, 22GDB041, 21LDB937, 21LDB416, 21JDB262, 21IDB491, 21IDA193, 21GDB207, and 21FDC754; (3) Kit ACC10540 - Kit UDI-DI: 10193489977547; Case UDI-DI 40193489977548; lot number: 23ADB410; (4) Kit ACC10540A - Kit UDI-DI 10195327330088; Case UDI-DI 40195327330089; lot numbers: 23HDA588, 23GDB107, and 23BDA684; (5) Kit ACC010717 - Kit UDI-DI 10195327458997; Case UDI-DI 40195327458998; lot numbers: 24ADA577, 23LDA731, and 23IDA629; (6) Kit DYNDJ1132 - Kit UDI-DI 10195327454821; Case UDI-DI 40195327454822; lot numbers: 23LDA733, 23LDA617, 23JDC021, and 23HDA569; (7) Kit DYNDJ1133 - Kit UDI-DI 10195327428532; Case UDI-DI 40195327428533; lot numbers: 24BMB124, 23LMB682, 23HMF259, and 23HMD765; (8) Kit DYNJ909501 - Kit UDI-DI 10195327384852; Case UDI-DI 40195327384853; lot number: 23GBI504; (9) Kit DYNJ909501A - Kit UDI-DI 10195327549817; Case UDI-DI 40195327549818; lot numbers: 24BBD129, 24ABK281, and 23KBR773; (10) Kit DYNJAA269 - Kit UDI-DI 10195327209049; Case UDI-DI 40195327209040; lot numbers: 24BBE224, 24ABM718, 24ABD835, 23LBB055, 23JBP694, 23JBQ783, 23IBU821, 23IBF348, 23HBP188, 23CBX318, 23CBX319, 23BBR494, 23ABI514,22LBI988, 22KBH009, and 22JBH478.