Z-1894-2016 Class I Terminated

Recalled by Carefusion 211 Inc dba Carefusion — Yorba Linda, CA

Recall Details

Product Type
Devices
Report Date
June 22, 2016
Initiation Date
May 17, 2016
Termination Date
September 28, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
922 units total (501 units in US)

Product Description

TCA Board, 16542A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.

Reason for Recall

CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).

Distribution Pattern

Worldwide Distribution - US (Nationwide) and the countries of Australia, Hungary, Paraguay, Belgium, Indonesia, Philippines, Bolivia, Ireland, Poland, Bermuda, India, Saudi Arabia, Canada, Japan, Singapore, Chile, Jordon, South Africa, China, Kuwait, Taiwan, Ecuador, Malaysia, Thailand, Egypt, Mexico, Turkey, France, Netherlands, United Kingdom, Guatemala, and Nicaragua.