Z-1894-2016 Class I Terminated

Recalled by Carefusion 211 Inc Dba Carefusion — Yorba Linda, CA

FDA device recall Z-1894-2016 was initiated by Carefusion 211 Inc Dba Carefusion on May 17, 2016 and is designated Class I. Reason for recall: CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing… The recall status is terminated (terminated Sep 28, 2017). Affected quantity: 922 units total (501 units in US).

Recall details

Product Type
Devices
Report Date
Jun 22, 2016
Initiation Date
May 17, 2016
Termination Date
Sep 28, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
922 units total (501 units in US)

Product description

TCA Board, 16542A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.

Reason for recall

CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).

Distribution pattern

Worldwide Distribution - US (Nationwide) and the countries of Australia, Hungary, Paraguay, Belgium, Indonesia, Philippines, Bolivia, Ireland, Poland, Bermuda, India, Saudi Arabia, Canada, Japan, Singapore, Chile, Jordon, South Africa, China, Kuwait, Taiwan, Ecuador, Malaysia, Thailand, Egypt, Mexico, Turkey, France, Netherlands, United Kingdom, Guatemala, and Nicaragua.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1894-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Carefusion 211 Inc Dba Carefusion? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1894-2016. Report a data issue: hello@regdatalab.com.