Z-1901-2025 Class II Ongoing
FDA device recall Z-1901-2025 was initiated by Beckman Coulter, Inc. on May 5, 2025 and is designated Class II. Reason for recall: Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could… The recall status is ongoing. Affected quantity: 70 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 11, 2025
- Initiation Date
- May 5, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 70 units
Product description
Access Testosterone assay, Catalog Number 33560
Reason for recall
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
Distribution pattern
US State: FL. UAE, Vietnam
Code information
UDI-DI: 15099590231347; Serial Number: 538081
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1901-2025
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026