Z-1901-2025 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

Recall Details

Product Type
Devices
Report Date
June 11, 2025
Initiation Date
May 5, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
70 units

Product Description

Access Testosterone assay, Catalog Number 33560

Reason for Recall

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

Distribution Pattern

US State: FL. UAE, Vietnam

Code Information

UDI-DI: 15099590231347; Serial Number: 538081