Z-1901-2025 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

FDA device recall Z-1901-2025 was initiated by Beckman Coulter, Inc. on May 5, 2025 and is designated Class II. Reason for recall: Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could… The recall status is ongoing. Affected quantity: 70 units.

Recall details

Product Type
Devices
Report Date
Jun 11, 2025
Initiation Date
May 5, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
70 units

Product description

Access Testosterone assay, Catalog Number 33560

Reason for recall

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

Distribution pattern

US State: FL. UAE, Vietnam

Code information

UDI-DI: 15099590231347; Serial Number: 538081

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1901-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1901-2025. Report a data issue: hello@regdatalab.com.