Z-1909-2025 Class II Ongoing

Recalled by Healthmark Industries Co., Inc. — Fraser, MI

FDA device recall Z-1909-2025 was initiated by Healthmark Industries Co., Inc. on May 2, 2025 and is designated Class II. Reason for recall: During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation… The recall status is ongoing. Affected quantity: 250000 units.

Recall details

Product Type
Devices
Report Date
Jun 11, 2025
Initiation Date
May 2, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
250000 units

Product description

Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010

Reason for recall

During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.

Distribution pattern

US Nationwide and Internationally to countries of: Canada, Malaysia.

Code information

Lot Numbers: 200001 200002 200003 200004 200005 200006 200007 200008 200009 200010 200011 200012 200013

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1909-2025
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1909-2025. Report a data issue: hello@regdatalab.com.