Z-1918-2015 Class II Terminated

Recalled by New Star Lasers, Inc. — Roseville, CA

FDA device recall Z-1918-2015 was initiated by New Star Lasers, Inc. on May 13, 2015 and is designated Class II. Reason for recall: During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially result in a discrepancy in the energy delivered (higher or lowe… The recall status is terminated (terminated October 28, 2016). Affected quantity: 52 units.

Recall Details

Product Type
Devices
Report Date
July 8, 2015
Initiation Date
May 13, 2015
Termination Date
October 28, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
52 units

Product Description

StoneLight 30 Laser System (NS3000). For use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, and General Surgery.

Reason for Recall

During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially result in a discrepancy in the energy delivered (higher or lower) from the value displayed on the Graphical User Interface display.

Distribution Pattern

Worldwide Distribution -- US, including the states of CA, IL, AZ, FL, OH, MA, MI, and AL; and the country of the Netherlands.

Code Information

Part Number: 0010-9400. Serial numbers: AMVM01, AMVM06, AMVM07, AMVG01, AMVP05, AMVP07, AMVP13, AMVS05, AMVK02, AMVS03, AMVS07, AMVE01, AMVS06, AMVP03, AMVS08. AMVS09, AMVS10, AMVS11, AMVS12, AMVM04, AMVM08, AMVM09, AMVM05, AMVS02, AMVK05, AMVP10, AMVS11, AMVP12, AMVP14, AMVP15, AMVP16, AMVP17, AMVP18. AMVS01, AMVK01, AMVK03, AMVK04, AMVK06, AMVK07, AMVK08, AMVK09, AMVK10, AMVM02, AMVM03, AMVM10, AMVP01, AMVP02, AMVP04, AMVP06, AMVP08, AMVP09, AMVS04.