Z-1922-2025 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1922-2025 was initiated by Olympus Corporation Of The Americas on May 6, 2025 and is designated Class II. Reason for recall: Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to… The recall status is ongoing. Affected quantity: 633,871 units (384146 US, 249725 OUS).

Recall details

Product Type
Devices
Report Date
Jun 18, 2025
Initiation Date
May 6, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
633,871 units (384146 US, 249725 OUS)

Product description

Brand Name: Powered Laser Surgical Instrument Product Name: Soltive SuperPulsed Laser Fibers - Single Use Model/Catalog Numbers: TFL-FBX150BS, TFL-FBX200BS, TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TLF-FBX940S, TFL-FBX150BS-JA (Japan Only), TFL-FBX550S-JA (Japan Only), TFL-FBX940S-JA (Japan Only), TFL-FBX200BS-JA (Japan Only), TFL-FBX150S-JA (Japan Only), TFL-FBX200S-JA (Japan Only), TFL-FBX365S-JA (Japan Only); Product Description: SOLTIVE Laser Fibers are delivery devices that transmit laser energy from the laser console to the treatment site through the fiber tip. The laser connector secures the device to the laser.

Reason for recall

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

Code information

Lot Code: All lot numbers for the following products: TFL-FBX150BS UDI: 00821925043923; TFL-FBX200BS UDI: 00821925043985; TFL-FBX150S UDI: 00821925043886; TFL-FBX200S UDI: 00821925043947; TFL-FBX365S UDI: 00821925044005; TFL-FBX550S UDI: 00821925044043; TFL-FBX940S UDI: 00821925044081; TFL-FBX150BS-JA (Japan Only); TFL-FBX550S-JA (Japan Only); TFL-FBX940S-JA (Japan Only); TFL-FBX200BS-JA (Japan Only); TFL-FBX150S-JA (Japan Only); TFL-FBX200S-JA (Japan Only); TFL-FBX365S-JA (Japan Only);

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1922-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1922-2025. Report a data issue: hello@regdatalab.com.