Z-1928-2019 Class III Terminated

Recalled by Roche Diagnostics Corporation — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
July 10, 2019
Initiation Date
April 4, 2019
Termination Date
May 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9 units

Product Description

cobas p 701 post-analytical units, Part Number 05349966001. For diagnostic sample storage and retrieval.

Reason for Recall

There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.

Distribution Pattern

US Nationwide Distribution in the states of AL, CA, FL, IA, IL, KY, MD, OR & PA

Code Information

Serial Numbers - 1001 through 1037, and 502