Z-1929-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 30, 2018
- Initiation Date
- March 30, 2018
- Termination Date
- November 29, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,052 units total
Product Description
Edwards Lifesciences' IntraClude Intra-Aortic Occlusion Device. Packaged in a sterile barrier system. Model Number: ICF100. Designed to occlude the ascending aorta.
Reason for Recall
Possible leak at distal open end of the strain relief, which causes medical staff to continue adding inflation fluid to maintain balloon pressure for adequate occlusion of the ascending aorta.
Distribution Pattern
Worldwide Distribution -- USA, to the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, MA, MI, MS, MO, NV, NJ, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, and WV; and, the countries of to Australia, Austria, Belgium, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Luxembourg, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Code Information
Lot No. 60876468, 60850885, 6060850886, 60850888, 60850889, 60850890, 60887217, 60887218, 60887219, 60887220, 60930494, 60934623, 60934624, 60934625, 60972889, 60972890, 60972892, 61014406, 61014408, 61014410, 61014411, 61050171, 61050172, 61078031, 61078032, 61097633, 61139239, 61139240, 61139241, 61139242, 61184964, 61184965, 67784966, 61184967, 61204586, 61204587, 61259626, 61259627, 61259628, and 61259629.