Z-1941-2014 Class II Terminated

Recalled by Meridian Bioscience Inc — Cincinnati, OH

FDA device recall Z-1941-2014 was initiated by Meridian Bioscience Inc on Apr 22, 2014 and is designated Class II. Reason for recall: Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration. The recall status is terminated (terminated Jun 30, 2014). Affected quantity: 35 instruments.

Recall details

Product Type
Devices
Report Date
Jul 9, 2014
Initiation Date
Apr 22, 2014
Termination Date
Jun 30, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35 instruments

Product description

illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.

Reason for recall

Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CO, FL, IA, IN, LA, MA, ME, MI, MN, NJ, NY, OH, PA, and WV. Internationally to Australia.

Code information

Serial Numbers: IP2669, IP2670, IP2671, IP2672, IP2673, IP2674, IP2675, IP2676, IP2677, IP2678, IP2680, IP2682, IP2683, IP2684, IP2685, IP2686, IP2687, IP2688, IP2689, IP2690, IP2691, IP2692, IP2693, IP2694, IP2695, IP2696, IP2697, IP2698, IP2699, IP2700, IP2701, IP2702, IP2703, IP2704, IP2705, IP2706.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1941-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Meridian Bioscience Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1941-2014. Report a data issue: hello@regdatalab.com.