Z-1949-2013 Class II Terminated
Recalled by Veridex, LLC
— Raritan, NJ
FDA device recall Z-1949-2013 was initiated by Veridex, LLC on Dec 28, 2012 and is designated Class II. Reason for recall: Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of… The recall status is terminated (terminated Aug 11, 2015). Affected quantity: US: 82 Units; Ex-US 78 Units.
Recall details
- Product Type
- Devices
- Report Date
- Aug 21, 2013
- Initiation Date
- Dec 28, 2012
- Termination Date
- Aug 11, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- US: 82 Units; Ex-US 78 Units.
Product description
CELLTRACKS AUTOPREP System (Software Version 2.4.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks AutoPrep System software 2.4.1 release notes Pub. No. J40135EN issued 2012-03-26 page 2, the document states that a new feature unique to software 2.4.1 allows the used to modify a batch in a new way. The user has additional optional fields in the edit sample information dialog box that allows one to enter expanded sample and patient information. The user can also edit the patient management setting for selected samples if not all samples have the same designation.
Reason for recall
Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of the issue and provide instructions to help prevent an occurrence on your CELLTRACKS AUTOPREP System and CELLTRACKS ANALYZER II.
Distribution pattern
Worldwide Distribution - USA Nationwide and the countries of Australia, Canada, China, India, Japan, Singapore, England, France, Germany, Italy and Spain.
Code information
CELLTRACKS AUTOPREP System (Software Version 2.4.1 & Below). Product Code: CT4297B (9541).
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1949-2013
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Veridex, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works