Z-1949-2013 Class II Terminated

Recalled by Veridex, LLC — Raritan, NJ

FDA device recall Z-1949-2013 was initiated by Veridex, LLC on Dec 28, 2012 and is designated Class II. Reason for recall: Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of… The recall status is terminated (terminated Aug 11, 2015). Affected quantity: US: 82 Units; Ex-US 78 Units.

Recall details

Product Type
Devices
Report Date
Aug 21, 2013
Initiation Date
Dec 28, 2012
Termination Date
Aug 11, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US: 82 Units; Ex-US 78 Units.

Product description

CELLTRACKS AUTOPREP System (Software Version 2.4.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks AutoPrep System software 2.4.1 release notes Pub. No. J40135EN issued 2012-03-26 page 2, the document states that a new feature unique to software 2.4.1 allows the used to modify a batch in a new way. The user has additional optional fields in the edit sample information dialog box that allows one to enter expanded sample and patient information. The user can also edit the patient management setting for selected samples if not all samples have the same designation.

Reason for recall

Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of the issue and provide instructions to help prevent an occurrence on your CELLTRACKS AUTOPREP System and CELLTRACKS ANALYZER II.

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Australia, Canada, China, India, Japan, Singapore, England, France, Germany, Italy and Spain.

Code information

CELLTRACKS AUTOPREP System (Software Version 2.4.1 & Below). Product Code: CT4297B (9541).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1949-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Veridex, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1949-2013. Report a data issue: hello@regdatalab.com.