Z-1950-2015 Class I Terminated

Recalled by Covidien LP (formerly Nellcor Puritan Bennett Inc.) — Boulder, CO

FDA device recall Z-1950-2015 was initiated by Covidien LP (formerly Nellcor Puritan Bennett Inc.) on May 8, 2015 and is designated Class I. Reason for recall: Reports where patients who recently switched from the current Shiley Neonatal and Pediatric products to the affected products experienced discomfort and in some cases, breathing difficulties, correspo… The recall status is terminated (terminated April 12, 2017). Affected quantity: 69,461 units total.

Recall Details

Product Type
Devices
Report Date
July 15, 2015
Initiation Date
May 8, 2015
Termination Date
April 12, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
69,461 units total

Product Description

Shiley Neonatal, Pediatric and Long Pediatric Tracheostomy Tube, Cuffless. Model Numbers: 2.5NEF,3.0NEF,3.5NEF,4.0NEF,4.5NEF,2.5PEF,3.0PEF,3.5PEF,4.0PEF,4.5PEF,5.0PEF,5.5PEF,5.0PELF,5.5PELF,6.0PELF,6.5PELF,2.5NCF,3.0NCF,3.5NCF,4.0NCF,4.5NCF,2.5PCF,3.0PCF,3.5PCF,4.0PCF,4.5PCF,5.0PCF,5.5PCF,5.0PLCF,5.5PLCF,6.0PLCF,6.5PLCF,3.0NEF-P,3.5NEF-P,4.0NEF-P,4.5NEF-P,3.0PEF-P,3.5PEF-P,4.0PEF-P,4.5PEF-P,5.0PEF-P,5.5PEF-P,

Reason for Recall

Reports where patients who recently switched from the current Shiley Neonatal and Pediatric products to the affected products experienced discomfort and in some cases, breathing difficulties, corresponding to a negative effect on oxygen levels.

Distribution Pattern

Worldwide Distribution. US nationwide, Canada, Australia, Belgium, Chile, Germany, Israel, Italy, Japan, Saudi Arabia, Singapore, South Africa, Turkey, and Uruguay.

Code Information

All products manufactured since 2/2012. Including All lot numbers beginning with 12, 13 and 14.