Z-1958-2014 Class II Terminated

Recalled by Focus Diagnostics Inc — Cypress, CA

Recall Details

Product Type
Devices
Report Date
July 9, 2014
Initiation Date
May 15, 2014
Termination Date
December 31, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
614 kits

Product Description

Simplexa" Flu A/B & RSV Direct, Model Number: MOL2650. 510(k) K120413  Simplexa" Flu A/B & RSV Direct. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is intended for use on the 3M Integrated Cycler instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory tract infection in conjunction with clinical and epidemiological risk factors. This test is intended for use as an aid in the differential diagnosis of influenza A, influenza B, and RSV viral infections in humans and is not intended to detect influenza C.

Reason for Recall

Focus Diagnostics is providing an urgent safety notice for a correction to the labeling for Simplexa" Flu A/B & RSV Direct (MOL2650). Focus Diagnostics received some customer complaints of Simplexa Flu A/B & RSV Direct assays with sporadic false signals, which may result in a higher false results rate due to a potential software spectral matrix and Direct Amplification Disc storage.

Distribution Pattern

Nationwide Distribution

Code Information

Model Number: MOL2650. All Lot Numbers. The expiration date of the 614 kits of MOL2650 that are impacted by this Correction ranges from 10/31/2014 to 5/31/2015.