Z-1976-2016 Class II Terminated

Recalled by Navilyst Medical, Inc., an AngioDyamics Company — Glens Falls, NY

Recall Details

Product Type
Devices
Report Date
June 22, 2016
Initiation Date
March 14, 2016
Termination Date
June 11, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
55 units

Product Description

Xcela PICC with PASV Valve Technology, Intermediate MST-45 Kit, 6F-55cm, UPN H965457440 Rx ONLY The Xcela PICC with PASV, Xcela Hybrid PICC with PASV, BioFlo PICC with PASV and BioFlo Hybrid PICC with PASV are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients; the sampling of blood; and for power injection of contrast media. Non-valve lumens are indicated for central venous pressure monitoring

Reason for Recall

PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.

Distribution Pattern

Worldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.

Code Information

Batch/Lots: 4718004, 4783051, 4788340 & 4796128. Use By Date Range 2016-03-31 to 2016-08-31