Z-2012-2019 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-2012-2019 was initiated by Siemens Medical Solutions USA, Inc on June 12, 2019 and is designated Class II. Reason for recall: A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X… The recall status is terminated (terminated September 18, 2020). Affected quantity: 5.

Recall Details

Product Type
Devices
Report Date
July 31, 2019
Initiation Date
June 12, 2019
Termination Date
September 18, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

AXIOM Artis FC, Model Number 5904433

Reason for Recall

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Distribution Pattern

US Nationwide Distribution

Code Information

Serial Numbers: 10436 10444 10425 10433 10435