Z-2023-2023 Class II Ongoing

Recalled by Luminex Corporation — Northbrook, IL

FDA device recall Z-2023-2023 was initiated by Luminex Corporation on Jun 5, 2023 and is designated Class II. Reason for recall: It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a… The recall status is ongoing. Affected quantity: 3987 units.

Recall details

Product Type
Devices
Report Date
Jul 5, 2023
Initiation Date
Jun 5, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3987 units

Product description

Verigene Enteric Pathogens Nucleic Acid Test (EP), REF 30-002-023

Reason for recall

It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.

Distribution pattern

Worldwide and US Nationwide Distribution

Code information

UDI/DI 00840487101568, Lot Numbers: 071322023D, 071822023B, 071822023C, 071922023E, 072022023D, 072622023D, 082422023G, 083022023E, 083122023E, 090622023A, 090822023D, 091322023D, 091422023D, 091422023E, 092022023E, 092122023D, 092722023D, 092822023D, 100422023D, 100722023E, 101422023E, 101822023D, 101822023E, 101922023D, 102022023D, 121322023D, 122722023C, 122722023D, 010323023D, 011323023G, 021323023A, 021323023B, 021523023D, 022123023E, 022323023D, 030123023F, 030223023D, 030223023E, 030623023A, 030823023D, 031623023D

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2023-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Luminex Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2023-2023. Report a data issue: hello@regdatalab.com.