Z-2081-2019 Class II Terminated

Recalled by Xcision Medical Systems, LLC — Columbia, MD

FDA device recall Z-2081-2019 was initiated by Xcision Medical Systems, LLC on Jul 3, 2019 and is designated Class II. Reason for recall: One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output. The recall status is terminated (terminated Jul 1, 2021). Affected quantity: 2.

Recall details

Product Type
Devices
Report Date
Jul 31, 2019
Initiation Date
Jul 3, 2019
Termination Date
Jul 1, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product description

GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.

Reason for recall

One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.

Distribution pattern

The products were distributed to the following US states: MD and TX.

Code information

Serial Numbers GP0002, GP0004; UDI: (01) 00850002615005

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2081-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Xcision Medical Systems, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2081-2019. Report a data issue: hello@regdatalab.com.