Z-2081-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 31, 2019
- Initiation Date
- July 3, 2019
- Termination Date
- July 1, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2
Product Description
GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.
Reason for Recall
One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.
Distribution Pattern
The products were distributed to the following US states: MD and TX.
Code Information
Serial Numbers GP0002, GP0004; UDI: (01) 00850002615005