Z-2082-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 22, 2015
- Initiation Date
- December 11, 2014
- Termination Date
- July 11, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 225 units
Product Description
Autopen Classic Automatic insulin delivery pen, 3.0ml x 2 unit 2-42 units, AN3800, Lot 7JH, NDC 08470-3800-01, The product is a non-sterile automatic subcutaneous injection device which delivers insulin.
Reason for Recall
The dials in the Autopens are not working - the dial does not stay dialed, but springs right back to 0. As a result, the patient could not dial up a dose of medicine to deliver via an injection.
Distribution Pattern
Nationwide Distribution including FL, IL, NV, NJ, OH, and TN.
Code Information
Lot 7JH, UPC Code 384703800014