Z-2110-2020 Class II Terminated

Recalled by The Binding Site Group, Ltd. — Birmingham, N/A

FDA device recall Z-2110-2020 was initiated by The Binding Site Group, Ltd. on Oct 7, 2019 and is designated Class II. Reason for recall: Assigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the reference… The recall status is terminated (terminated Jan 11, 2022). Affected quantity: 478 kits.

Recall details

Product Type
Devices
Report Date
Jun 3, 2020
Initiation Date
Oct 7, 2019
Termination Date
Jan 11, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
478 kits

Product description

Optilite Optilite Rheumatoid Factor Kit, Ref: LK151.OPT.A, IVD, Rx Only, UDI: 05051700020015 (kit ea. containing four components)

Reason for recall

Assigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the reference material RF serum. This could result in reporting abnormal (false positive) results, which may lead to unnecessary additional serological testing and a delay in treatment.

Distribution pattern

US: AL, AR, AZ, FL, GA, KY, MA, MD, MI, MN, MS, NC, NE, NY, OR, PA, TX, VT, WA, WI OUS: None

Code information

Kit Lot Number: 428224

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2110-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for The Binding Site Group, Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2110-2020. Report a data issue: hello@regdatalab.com.