Z-2110-2020 Class II Terminated

Recalled by The Binding Site Group, Ltd. — Birmingham, N/A

Recall Details

Product Type
Devices
Report Date
June 3, 2020
Initiation Date
October 7, 2019
Termination Date
January 11, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
478 kits

Product Description

Optilite Optilite Rheumatoid Factor Kit, Ref: LK151.OPT.A, IVD, Rx Only, UDI: 05051700020015 (kit ea. containing four components)

Reason for Recall

Assigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the reference material RF serum. This could result in reporting abnormal (false positive) results, which may lead to unnecessary additional serological testing and a delay in treatment.

Distribution Pattern

US: AL, AR, AZ, FL, GA, KY, MA, MD, MI, MN, MS, NC, NE, NY, OR, PA, TX, VT, WA, WI OUS: None

Code Information

Kit Lot Number: 428224