Z-2123-2019 Class II Terminated

Recalled by Burlington Medical, LLC — Newport News, VA

Recall Details

Product Type
Devices
Report Date
August 7, 2019
Initiation Date
June 7, 2019
Termination Date
April 28, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
39

Product Description

Sleeve, Product Number Sleeve, Part Numbers RESL, SL 3 L, SL 3 M, SL 3 S, L 3 X, SL 3 XS, UESL Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Reason for Recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Distribution Pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

Code Information

Serial Numbers: 444749 418756 418757 408712 388138 388139 407957 339006 419340 391212 381909 381862 391205 414767 419574 419561 419562 419190 419189 391214 443199 418796 350481 407814 407813 443198 443200 443201 443197 391203 381803 379683 419191 419192 419107 383262 383261 387157 387158