Z-2123-2025 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, Inc. — Rochester, NY

Recall Details

Product Type
Devices
Report Date
July 23, 2025
Initiation Date
June 10, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
414,000 (OUS)

Product Description

VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);

Reason for Recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Distribution Pattern

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.

Code Information

Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides); UDI-DI: (1) 10758750030293, (2) 10758750030309; Lot Numbers: All lots within expiry.