Z-2133-2025 Class II Ongoing

Recalled by GE Medical Systems, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
July 30, 2025
Initiation Date
May 23, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

SIGNA PET/MR

Reason for Recall

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

Distribution Pattern

Worldwide distribution - US Nationwide and the country of China.

Code Information

GTIN: 00840682105378