Z-2147-2024 Class II Ongoing

Recalled by Ethicon, Inc. — Raritan, NJ

Recall Details

Product Type
Devices
Report Date
June 26, 2024
Initiation Date
May 10, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
45828 units

Product Description

PDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in ophthalmic surgery. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable. PRODUCT CODE: PDP304H PDP316H PDP341H

Reason for Recall

Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

Distribution Pattern

Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.

Code Information

GTIN: 10705031047976 10705031048027 10705031048119 PRODUCT CODE PRODUCT LOTS PDP304H TLMJLZ PDP316H UAMKDC PDP341H UBMCZA