Z-2152-2017 Class II Terminated

Recalled by Hans Pausch Rontgengeratebau Gmbh — Erlangen

FDA device recall Z-2152-2017 was initiated by Hans Pausch Rontgengeratebau Gmbh on Mar 14, 2017 and is designated Class II. Reason for recall: During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological… The recall status is terminated (terminated Jun 14, 2019). Affected quantity: 10 systems.

Recall details

Product Type
Devices
Report Date
Jun 7, 2017
Initiation Date
Mar 14, 2017
Termination Date
Jun 14, 2019
Voluntary/Mandated
FDA Mandated
Product Quantity
10 systems

Product description

Pausch Uroview FD fluoroscopic X-ray system UDI number (01)04260491620040 Primarily for urological applications (functional X-ray diagnostics, endourology and minimal invasive urology / surgery). The system, which includes a radiologic / urologic treatment table, maybe used for urological treatment, planning and diagnostic procedures, including but not limited to.

Reason for recall

During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if footswitch + exposure + first primary barrier are closed the signal will be transmitted to the generator for activating X-ray. During intended use only with this footswitch X-ray should be activated.

Distribution pattern

Nationwide Distribution

Code information

model numbers: 01169033 and 01169034 with the serial numbers 0103-0112

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2152-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Hans Pausch Rontgengeratebau Gmbh? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2152-2017. Report a data issue: hello@regdatalab.com.