Z-2152-2017 Class II Terminated

Recalled by Hans Pausch Rontgengeratebau Gmbh — Erlangen

Recall Details

Product Type
Devices
Report Date
June 7, 2017
Initiation Date
March 14, 2017
Termination Date
June 14, 2019
Voluntary/Mandated
FDA Mandated
Product Quantity
10 systems

Product Description

Pausch Uroview FD fluoroscopic X-ray system UDI number (01)04260491620040 Primarily for urological applications (functional X-ray diagnostics, endourology and minimal invasive urology / surgery). The system, which includes a radiologic / urologic treatment table, maybe used for urological treatment, planning and diagnostic procedures, including but not limited to.

Reason for Recall

During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if footswitch + exposure + first primary barrier are closed the signal will be transmitted to the generator for activating X-ray. During intended use only with this footswitch X-ray should be activated.

Distribution Pattern

Nationwide Distribution

Code Information

model numbers: 01169033 and 01169034 with the serial numbers 0103-0112