Z-2153-2012 Class II Terminated

Recalled by Varian Medical Systems, Inc. Oncology Systems — Palo Alto, CA

Recall Details

Product Type
Devices
Report Date
August 15, 2012
Initiation Date
July 2, 2012
Termination Date
July 1, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 Units.

Product Description

Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA also includes optional tools to manage the oncology department's clinical and business information.

Reason for Recall

A software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported.

Distribution Pattern

Nationwide Distribution including the states of New York and Ohio.

Code Information

Serial Numbers/CODES: HIT0535, HIT2953