Z-2175-2024 Class II Ongoing
Recalled by Maquet Cardiovascular, LLC
— Wayne, NJ
FDA device recall Z-2175-2024 was initiated by Maquet Cardiovascular, LLC on May 17, 2024 and is designated Class II. Reason for recall: Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended. The recall status is ongoing. Affected quantity: 430,037 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 3, 2024
- Initiation Date
- May 17, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 430,037 units
Product description
The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2, Model Number: VH-4000. 2. VasoView Hemopro 2 with Vasoshield. Model Number: VH-4001. The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
Reason for recall
Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.
Code information
1. Model Number: VH-4000. UDI-DI: 00607567700406. All lot numbers within expiration. 2. Model Number: VH-4001. UDI-DI: 00607567700451. All lot numbers within expiration.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2175-2024
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Maquet Cardiovascular, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works