Z-2175-2025 Class II Ongoing

Recalled by Remote Diagnostic Technologies Ltd. — Farnborough

FDA device recall Z-2175-2025 was initiated by Remote Diagnostic Technologies Ltd. on May 30, 2025 and is designated Class II. Reason for recall: Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall)… The recall status is ongoing. Affected quantity: 55.

Recall details

Product Type
Devices
Report Date
Aug 6, 2025
Initiation Date
May 30, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
55

Product description

Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R

Reason for recall

Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.

Distribution pattern

Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.

Code information

REF/UDI-DI: 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932. Software Version: vx.28

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2175-2025
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2175-2025. Report a data issue: hello@regdatalab.com.